VEOLIA Sievers M500 Online TOC Analyzer

Detect critical organic contaminants with unamatched reliability, robustness, and accuracy in pharmaceutical Purified Water (PW) and Water for Injection (WFI) applications with the new Sievers M500 Online TOC Analyzer The Sievers M500 is our third generation online TOC analyzer for pharmaceutical grade water monitoring designed for accuracy, efficiency, and integrity. Building upon the proven results of the 500 RL, the M500 improves our industry-leading performance and adds a suite of cutting-edge features: Super iOS 4-port sampler automates grab samples to increase efficiency. Reducing analysis time by 50% facilitates utilization of real-time data, early detection, and process control. 10-inch touchscreen enables faster and easier setup and operation. Standardized and customized protocols improve productivity. Enhanced data features improve data integrity.

VEOLIA Sievers M500 Online TOC Analyzer
  • VEOLIA Sievers M500 Online TOC Analyzer
0
*We exclusively distribute our products in mainland China. Should you be interested in discussing potential agency opportunities, please reach out via the provided email address.
E-mail:samwu@sun-techcn.com
  • Product details
  • Specification parameters
  • Detect critical organic contaminants with unamatched reliability, robustness, and accuracy in pharmaceutical Purified Water (PW) and Water for Injection (WFI) applications with the new Sievers M500 Online TOC Analyzer

    The Sievers M500 is our third generation online TOC analyzer for pharmaceutical grade water monitoring designed for accuracy, efficiency, and integrity. Building upon the proven results of the 500 RL, the M500 improves our industry-leading performance and adds a suite of cutting-edge features:

    • Super iOS 4-port sampler automates grab samples to increase efficiency.

    • Reducing analysis time by 50% facilitates utilization of real-time data, early detection, and process control.

    • 10-inch touchscreen enables faster and easier setup and operation.

    • Standardized and customized protocols improve productivity.

    • Enhanced data features improve data integrity.

    M500 Online Features & Benefits

    Accuracy. Efficiency. Integrity.

    • 150% faster analysis time

    • Advanced communications: Ethernet, WiFi, 4-20mA, Modbus, and Serial connectivity

    • 10-inch color touchscreen

    • Super iOS 4-port sampler automates grab samples to improve productivity

    • Simultaneous TOC & conductivity measurement

    • Automated system suitability calibration, & validation

    • 50% faster validation protocols

    • IP55 & full IP connections

    • Same footprint as 500 RL for easy upgrades

    Data Integrity

    The Sievers M500 leverages comprehensive industry-leading TOC data management tools to ensure data is secure. The Sievers M500 supports compliance to 21 CFR Part 11 and adherence to US FDA and other Pharmacopoeia Data Integrity guidelines using new data transfer, data security and data management features.

    Data Transfer

    • Remote access using Ethernet and WiFi

    • Advanced communications using 4-20 mA, Modbus, and binary

    Data Security

    • Password protection

    • Data encryption

    • Customizable access, roles, and permissions

    Web-based Data Management

    • Closed system architecture

    • Customizable data transfer and export


    Detect critical organic contaminants with unamatched reliability, robustness, and accuracy in pharmaceutical Purified Water (PW) and Water for Injection (WFI) applications with the new Sievers M500 Online TOC Analyzer

    The Sievers M500 is our third generation online TOC analyzer for pharmaceutical grade water monitoring designed for accuracy, efficiency, and integrity. Building upon the proven results of the 500 RL, the M500 improves our industry-leading performance and adds a suite of cutting-edge features:

    • Super iOS 4-port sampler automates grab samples to increase efficiency.

    • Reducing analysis time by 50% facilitates utilization of real-time data, early detection, and process control.

    • 10-inch touchscreen enables faster and easier setup and operation.

    • Standardized and customized protocols improve productivity.

    • Enhanced data features improve data integrity.

    M500 Online Features & Benefits

    Accuracy. Efficiency. Integrity.

    • 150% faster analysis time

    • Advanced communications: Ethernet, WiFi, 4-20mA, Modbus, and Serial connectivity

    • 10-inch color touchscreen

    • Super iOS 4-port sampler automates grab samples to improve productivity

    • Simultaneous TOC & conductivity measurement

    • Automated system suitability calibration, & validation

    • 50% faster validation protocols

    • IP55 & full IP connections

    • Same footprint as 500 RL for easy upgrades

    Data Integrity

    The Sievers M500 leverages comprehensive industry-leading TOC data management tools to ensure data is secure. The Sievers M500 supports compliance to 21 CFR Part 11 and adherence to US FDA and other Pharmacopoeia Data Integrity guidelines using new data transfer, data security and data management features.

    Data Transfer

    • Remote access using Ethernet and WiFi

    • Advanced communications using 4-20 mA, Modbus, and binary

    Data Security

    • Password protection

    • Data encryption

    • Customizable access, roles, and permissions

    Web-based Data Management

    • Closed system architecture

    • Customizable data transfer and export


  • M500 Online Specifications

    Total Organic Carbon

    Linear Range0.03 to 2,500 ppb as TOC
    Accuracy± 5% of measurement; ±0.1 ppb
    Precision± 1% of measurement; ±0.3 ppb
    pH Range5.5 to 8.0 in order to meet accuracy specification
    Analysis ModesOn-line, on-line averaged, on-line timed, Grab
    Analysis Time1,23 minutes for continuous online measurements
    Ozone Compatibility50 ppb Ocontinuous; 200 ppb 03 for 2 hours daily
    External Flow RateMinimum 50 mL/min
    Sample Temperature1–95 °C (34–203 °F) (withstands short-term exposure)
    Sample PressureUp to 100 psig
    InterferencesInsensitive to organic heteroatoms
    Calibration StabilityTypically stable for 12 months
    Display Readout3 significant digits

    Conductivity

    Raw Conductivity Range0.01 to 800 µS/cm
    Conductivity Accuracy0.005 µS/cm or 1%, whichever is greater
    Conductivity Precision<1.0% RSD

    Instrument

    Power Requirements100–240 VAC, 70 W, 50/60 Hz
    FusesNo user-replaceable fuses
    Normal Operating EnvironmentIntended for indoor use only
    Ambient Temperature5–40°C (41–104°F)
    Maximum Relative HumidityUp to 95%, noncondensing
    Maximum Altitude3,000 m (9,843 ft)
    InputsTwo isolated binary inputs
    OutputsSerial (RS-232), one USB, three 40-20 mA, four alarms, one Ethernet
    Installation/OvervoltageII (protects against transients present in Category II power)
    Safety Certifications

    CE, ETL listed. Conforms to UL Std. 61010-1. Certified to CSA 22.2 No. 61010-1.

    Pollution Degree2 (normally only non-conductive pollution)
    DisplayBacklit 10.1", 1280x800, touchscreen display
    SizeH: 43.4 cm (17.1 in); W: 55.9 cm (22.0 in); D: 28.7 cm (11.3 in)
    Weight16.3 kg (36 lb)
    IP RatingIP 55
    Optional Wi-Fi802.11ac/a/b/g/n Dual-band 2.4/5 GHz
    Industrial Communications ProtocolModbus TCP/IP

     

    1Time to first measurement is 10 minutes

    2Base model analysis time is six minutes for continuous online measurement